A 51-year-old female with high myopia and bilateral cataracts presented with CDVA of 0.2 (OD) and 0.5 (OS). Ocular biometry was obtained with a new-generation 30KHz dynamic swept-source optical biometer Zalioth (TowardPi ).
Real-time dynamic panoramic axial OCT imaging ensured true visual axis measurement (fig.1)

Comprehensive topography enabled precise assessment of anterior, posterior, and total corneal astigmatism (fig.2).

Both eyes showed >0.75 D astigmatism, indicating toric IOL correction. Given >0.75 D of astigmatism in both eyes, toric IOL correction was indicated. 6-mm spherical aberration and 4-mm HOAs were evaluated with IOL Optimization function on device, Kappa angle and crystalline lens analysis as well (fig.3).

In the right eye, total corneal HOAs of 0.10 μm, Kappa angle of 0.117 mm, and pupil diameter of 3.27 mm met premium IOL criteria. IOL calculation using integrated Kane and EVO 2.0 formulas was cross-validated with Barrett Toric formula (fig.4).

A toric EDOF IOL (DFW300, SE +15.5 D, astigmatism +3.00 D) was implanted in the right eye. At one week, UDVA reached 1.2, with defocus curve confirming continuous clear vision across distance, intermediate, and near. The patient reported a high level of satisfaction.
Case courtesy: Dr. Jie Liu from Hebei Children’s Hospital.
*Zalioth is now CE-marked under EU MDR.